What is the minimum immersion depth with dry-well calibrators?
The minimum immersion depth for dry-well calibrators, essentially, depends on the axial gradient and the desired accuracy of the calibration. It is generally recommended to calibrate with the probe sitting on the bottom of the sleeve bore. For shor ...
moreHow must the inserts be designed in order to get the best results?
The inserts should be drilled in accordance with the diameter of the test items. In general, the bore diameter, taking the thermal expansion of the test item into account, should be selected to be as small as possible. Air gap between the thermomet ...
more